Safety alert issued over medical air connection

A National Patient Safety Alert has been issued asking all providers that use piped medical air to eliminate the risk of inadvertently connecting patients to medical air via a flowmeter instead of oxygen.

 

Air flowmeters attached to piped medical air outlets are primarily used to drive the administration of nebulised medication; typically for short periods to manage respiratory conditions. Most other uses of piped medical air do not require an air flowmeter.

Due to the proximity of the piped medical air and oxygen outlets at the bedside, and the similarity in design of flowmeters, there is a significant risk when using air flowmeters that patients may be inadvertently connected to medical air instead of oxygen.

The National Patient Safety Alert – Eliminating the risk of inadvertent connection to medical air via a flowmeter asks providers to purchase alternative devices that do not require medical air to be delivered via an air flowmeter. Following this, all medical air flowmeters except those tethered to equipment for niche use should be discarded, and all medical air outlets no longer required should be reversibly capped off. 

Failure to take the actions required under any National Patient Safety Alert may lead to CQC taking regulatory action.

 

Latest Issues

EBME Expo 2026

Coventry Building Society Arena, Coventry
24th – 25th June 2026

AfPP Regional Conferences: Manchester

INNSiDE by Meliá Manchester
20th June 2026

Endoscopic Anterior Skull Base Surgery: Hands-On Cadaveric Course

Division of Anatomy, University of Leeds
29th- 30th June 2026

BLOCKED – Advanced+ | The Wrightington Regional Anaesthesia Interest Group Cadaveric Course

Wrightington Conference Centre
Tuesday 7th – Wednesday 8th July 2026

AESCULAP ACADEMY LIVE - Circular Economy in Action

B. Braun Business Centre, Sheffield
Friday 10th July 2026

AfPP Regional Conferences: Bristol

BAWA Leisure
18th July 2026